Merlin II Protocol Analyzer User’s Manual

CATC

SW Version 2.50

Step 2

Enter MerlinII.exe -CLK

The application appears on the screen.

Step 3

Select Setup > Recording Options to open the Recording Options
dialog box.

Step 4

Set Inquiry Timeout to 0.

Step 5

Press the Start Recording button run inquiry for at least 5 minutes.

This will bring the analyzer to its operating temperature.

Step 6

Press Stop to stop the inquiry.

Step 7

Select Setup >Connectors from the menu.

The following window appears:

Step 8

Select the option Output Clock for Calibration.

At this point the clock signal should be directed to the 'TRIG OUT'
BNC connector. The frequency counter should start showing a value
around 26 MHz now.

Step 9

Click on the Clock Cal... button. Please note that this button is
visible only when the Merlin II application is run with the -CLK
argument.