Rosemount 5300 series – Emerson Process Management ROSEMOUNT 5300 User Manual
Page 96
Reference Manual
00809-0100-4530, Rev BA
July 2009
Rosemount 5300 Series
5-24
Rosemount Radar Master (RRM) includes tools to estimate the dielectric
constant of the current product:
•
The Dielectric Chart lists the dielectric constant of a large number of
products. The Dielectric Chart can be opened with one of the following
methods:
- Choose the View>Dielectric Constant Chart menu option
- Click the Dielectric Chart button
in the Configuration Wizard -
Environment window
- Choose the Setup>Tank menu option and click the Dielectric Chart
button
in the Tank/Environment window
•
The Dielectric Calculator
lets you calculate the dielectric constant
of the Upper Product based on the following input:
- actual upper product thickness,
- the dielectric constant value stored in the transmitter, and
- the upper product thickness presented by the transmitter.
The Dielectric Calculator is available via the Dielectric Calculator
button
in the Configuration Wizard - Environment window or the
Tank/Environment window.