Reference Manual

00809-0100-4530, Rev BA

July 2009

Rosemount 5300 Series

B-18

Figure B-20. Approval Label
IECEx Foundation Fieldbus
model

I7

Intrinsically Safe

Ex ia IIC T4 (T

amb



-50 °C to +60 °C).

IECEx NEM 06.0001X.

F

OUNDATION

Fieldbus: U

i

=30 V, I

i

=300 mA, P

i

=1.5 W, C

i

=0 nF,

L

i

=0 mH.

Installation Drawing: 9240 030-938.

Conditions of Certification

The intrinsically safe circuits do not withstand the 500V AC test as specified in
EN 50020 clause 6.4.12.

Probes covered with plastic and/or with plastic discs will have a
non-conducting area that exceeds the maximum permissible area for Group
IIC accoring to IEC 60079-01 clause 7.3: 20 cm

2

for Zone 1 and 4 cm

2

for

Zone 0. Therefore, when the probe is used in a potentially explosive
atmosphere, appropriate measures must be taken to prevent electrostatic
discharge.

Impact and friction hazards need then to be considered according to IEC
60079-0 clause 8.1.2 when the transmitter exposed to the exterior
atmosphere of the tank is made with light metal alloys and used in Zone 0.