Rosemount 5300 series – Emerson Process Management ROSEMOUNT 5300 User Manual
Page 229
Reference Manual
00809-0100-4530, Rev BA
July 2009
B-7
Rosemount 5300 Series
Figure B-5. Approval Label
ATEX Foundation Fieldbus
E1
The following information is provided as part of the label of the
transmitter:
•
Name and address of the manufacturer (Rosemount).
•
CE Conformity Marking
•
Complete model number
•
The serial number of the device
•
Year of construction
•
Marking for explosion protection:
•
EEx iad IIC T4 (-40
°
C a < +60 ° C) • Nemko ATEX certificate number: Nemko 04ATEX1073X • U m =250 VAC Special Conditions for Safe Use (X): The intrinsically safe circuits do not withstand the 500V AC test as specified in Probes covered with plastic and/or with plastic discs will have a 2 ). Therefore, when the probe is used in a potentially explosive atmosphere, Impact and friction hazards need then to be considered according to EN
EN 50020 clause 6.4.12
non-conducting area that exceeds the maximum permissible areas for Group
IIC and Category II 1G according to EN 50284 clause 4.4.3 (4 cm
appropriate measures must be taken to prevent electrostatic discharge.
50284 clause 4.3.1 when the transmitter exposed to the exterior atmosphere
of the tank is made with light metal alloys and used in category II 1 G.