Rosemount 5300 series – Emerson Process Management ROSEMOUNT 5300 User Manual
Page 54
Reference Manual
00809-0100-4530, Rev BA
July 2009
Rosemount 5300 Series
3-24
Solid Applications
Pull the probe rope through a
suitable anchoring point, e.g. a
welded eye and fasten it with two
clamps. It is recommended that the
probe is slack in order to prevent
high tensile loads.
The sag should be at least
1.5 in./10 ft (1 cm/m) of the probe
length.
Alternative chuck for Flexible
Single Lead probes
Loosen the screws. Pull the probe
rope through a suitable anchoring
point, e.g. a welded eye.
Tighten the screws. The required
torque and hex key dimensions:
4 mm wire: 15 Nm, 4 mm
6 mm wire: 25 Nm, 5 mm
1cm/m