beautypg.com

Rosemount 5300 series, Reference manual – Emerson Process Management ROSEMOUNT 5300 User Manual

Page 300

background image

Reference Manual

00809-0100-4530, Rev BA

July 2009

Rosemount 5300 Series

I-2

CYCLE_TYPE

19

Identifies the block execution methods available for this resource.

DD_RESOURCE

09

String identifying the tag of the resource which contains the Device Description for
this resource.

DD_REV

13

Revision of the DD associated with the resource - used by an interface device to locate
the DD file for the resource.

DEFINE_WRITE_LOCK

60

Allows the operator to select how WRITE_LOCK behaves. The initial value is “lock
everything”. If the value is set to “lock only physical device” then the resource and
transducer blocks of the device will be locked but changes to function blocks will be
allowed.

DETAILED_STATUS

55

Indicates the state of the transmitter. See Resource Block detailed status codes.

DEV_REV

12

Manufacturer revision number associated with the resource - used by an interface device
to locate the DD file for the resource.

DEV_STRING

43

This is used to load new licensing into the device. The value can be written but will
always read back with a value of 0.

DEV_TYPE

11

Manufacturer’s model number associated with the resource - used by interface devices
to locate the DD file for the resource.

DIAG_OPTION

46

Indicates which diagnostics licensing options are enabled.

DISTRIBUTOR

42

Reserved for use as distributor ID. No Foundation enumerations defined at this time.

DOWNLOAD_MODE

67

Gives access to the boot block code for over-the-wire downloads.
0 = Uninitialized
1 = Run mode
2 = Download mode

FAULT_STATE

28

Condition set by loss of communication to an output block, fault promoted to an output
block or physical contact. When FAIL_SAFE condition is set, then output function blocks
will perform their FAIL_SAFE actions.

FAILED_ACTIVE

72

Enumerated list of failure conditions within a device.

FAILED_ALM

73

Alarm indicating a failure within a device which makes the device non-operational.

FAILED_ENABLE

70

Enabled FAILED_ALM alarm conditions. Corresponds bit for bit to the FAILED_ACTIVE.
A bit on means that the corresponding alarm condition is enabled and will be detected. A
bit off means the corresponding alarm condition is disabled and will not be detected.

FAILED_MASK

71

Mask of FAILED_ALM. Corresponds bit of bit to FAILED_ACTIVE. A bit on means that
the condition is masked out from alarming.

FAILED_PRI

69

Designates the alarming priority of the FAILED_ALM.

FB_OPTION

45

Indicates which function block licensing options are enabled.

FEATURES

17

Used to show supported resource block options.
The supported features are: SOFT_WRITE_LOCK_SUPPORT,
HARD_WRITE_LOCK_SUPPORT, REPORTS, and UNICODE

FEATURES_SEL

18

Used to select resource block options.

FINAL_ASSY_NUM

54

The same final assembly number placed on the neck label.

FREE_SPACE

24

Percent of memory available for further configuration. Zero in a preconfigured device.

FREE_TIME

25

Percent of the block processing time that is free to process additional blocks.

GRANT_DENY

14

Options for controlling access of host computers and local control panels to operating,
tuning, and alarm parameters of the block. Not used by device.

HARD_TYPES

15

The types of hardware available as channel numbers.

HARDWARE_REV

52

Hardware revision of the hardware that has the resource block in it.

HEALTH_INDEX

84

Parameter representing the overall health of the device, 100 being perfect and 1 being
non-functioning. The value is based on the active PWA alarms.

ITK_VER

41

Major revision number of the inter operability test case used in certifying this device as
interoperable. The format and range are controlled by the Fieldbus Foundation.

LIM_NOTIFY

32

Maximum number of unconfirmed alert notify messages allowed.

MAINT_ACTIVE

77

Enumerated list of maintenance conditions within a device.

Parameter

Index

Number

Description