Reference Manual

00809-0100-4530, Rev BA
July 2009

3-17

Rosemount 5300 Series

Tri-Clamp Connection

Figure 3-13. Tri-Clamp tank
connection.

Tri-Clamp

Probe

Transmitter head

Gasket

Tank

1. Place a gasket on top of the tank flange.

2. Lower the transmitter and probe into the tank.

3. Fasten the Tri-Clamp to the tank with a clamp.

4. Loosen the nut that connects the transmitter

housing to the probe slightly.

5. Rotate the transmitter housing so the cable

entries/display face the desired direction.

6. Tighten the nut.

Nut