July 2009

Rosemount 5300 Series

7-34

The reference pulse is not detected.

• The tank is full
• The transmitter is configured with

wrong probe type

• Reference Amplitude Threshold is

incorrect

• Check the product level
• Check that correct probe type is configured
• Check Reference Amplitude Threshold

Incorrect Interface Level reading.

• Interface threshold incorrect
• Upper Product Dielectric constant

incorrect

• Adjust Interface threshold, see “Interface

Pulse not Found” on page 7-8

• Check the Upper Product Dielectric

constant, see “Basic Configuration
Parameters” on page 5-4

Incorrect level reading.

• Configuration error
• Disturbing objects in the tank

• Check the Tank Height parameter
• Check status information and diagnostic

information

• Check that the transmitter has not locked

on an interfering object, see “Disturbance
Echo Handling” on page 7-7

• Adjust the Surface amplitude threshold, see

“Surface Pulse Not Found” on page 7-5

Integral display does not work.

• Check the display configuration
• Check loop power
• Check Display connection
• Contact Emerson Process Management

Service Department

(1)

F

OUNDATION

Fieldbus Card to

Transmitter Communication Fault

• Verify Device Mode setting, should be

F

OUNDATION

Fieldbus

(Parameter: ENV_DEVICE_MODE)

• Restart method from Resource Block
• Reboot gauge (Cycle Power)

Level Measurement Failure

• Check Power Supply
• Check the gauge configuration (Transducer

Block)

• Check that the mechanical installation is

correct

Temperature Measurement Failure

• Check ambient temperature

(2)

• Restart gauge
• Contact Emerson Process Management

Service Department

Volume Measurement Failure

• Restart gauge
• Check gauge configuration using PC Based

configuration tool

No surface echo

• Check signal strength
• Restart transmitter
• See “Surface Pulse Not Found” on

page 7-5

DB Error/ Microwave Unit Error/
Configuration Error/ Other Error

• Restart transmitter
• Set database to default; load default

Database

• Download Application Software
• Call Service Center

SW Error/ Display Error/ Analog Out
Error

• Restart transmitter
• Call Service Center

(1) A malfunctioning display panel may only be replaced by service personnel at the Emerson Process Management Service Department. A display

must not be replaced when the transmitter is in operation.

(2) If the 5300 transmitter has been exposed to temperatures outside the specified limits, the device may stop its normal operation.

Symptom

Possible cause

Action