Rosemount 5300 series, Reference manual – Emerson Process Management ROSEMOUNT 5300 User Manual
Page 276
Reference Manual
00809-0100-4530, Rev BA
July 2009
Rosemount 5300 Series
E-2
Figure E-1. Range Values.
5. Perform a two-point calibration check of the transmitter by applying level
to two points on the probe within the measuring range. Verify that the
current output corresponds to the level input values using a known
reference measurement.
This step verifies that the analog output is correct in the operating range
and that the Primary Variable is properly configured.
Note that the applied level has to be between the Upper and Lower
Range values, otherwise the transmitter enters alarm mode. If the
applied level is outside the Maximum Measuring Range, the level
reading accuracy may be reduced. For best performance, use the
4-20 mA range points as calibration points. See Figure E-1.
6. Enable write protection.
375 HART Sequence: [3, 2, 1, 2, 1].
7. Restore the loop to full operation.
8. Remove the bypass from the safety PLC or otherwise restore normal
operation.
9. Document the test results for future reference.
4mA
20mA
Upper Transition Zone
Lower Transition Zone
R
a
ng
e 0 -100
%
Maximum
Measuring Range
Upper Range Value
Lower Range Value