Changing a probe -29, Probe and firmware compatibility -29, Rosemount 5300 series – Emerson Process Management ROSEMOUNT 5300 User Manual
Page 159: Reference manual, Changing a probe probe and firmware compatibility
Reference Manual
00809-0100-4530, Rev BA
July 2009
7-29
Rosemount 5300 Series
CHANGING A PROBE
Probe and Firmware
Compatibility
Transmitter heads with a firmware version earlier than 1.A4 (manufacturing
date before 2008-06-18) are not compatible with HP/HTHP/C probes marked
with R2.
Transmitter heads with a firmware version 1.A4 or later are compatible with
HP/HTHP probes without the R2 marking when the Trim Near Zone function
is performed, as illustrated below.
Only probes with the VC marking are compatible with the Dynamic Vapor
Compensation function. See Appendix C: Advanced Configuration to
determine if the Dynamic Vapor Compensation function is supported by the
transmitter head.
Table 7-1. Compatibility
between probe type and
firmware version
Compatibility Between Probe Types and Firmware Versions
Firmware Version
Probe Type
Standard
HP / HTHP without
R2 marking
HP / HTHP / C with
R2 marking only
(1)
HTHP with both
R2 and VC marking
Firmware version earlier
than 1.A4
Yes
Yes
No
No
Firmware version 1.A4
Yes
Yes
(2)
Yes
No
Firmware version 2.A2 or
later
Yes
Yes
Yes
(3)
Yes
(4)
(1) The R2 marking is on the housing seal or the adapter as seen in the figure.
(2) Trim Near Zone function is required.
(3) When Dynamic Vapor Compensation is not used.
(4) The probe requires that Dynamic Vapor Compensation is activated in the device for full functionality.
R 2
R 2
R 2
R 2
R2 VC
R2 VC
R2
R2