Errors -36, Rosemount 5300 series, Reference manual – Emerson Process Management ROSEMOUNT 5300 User Manual
Page 166: Errors
Reference Manual
00809-0100-4530, Rev BA
July 2009
Rosemount 5300 Series
7-36
Errors
Error messages that may be displayed on the Integral Display, on a 375
Handheld Communicator, in AMS, or in RRM are shown in Table 7-5. Errors
normally result in Analog Output alarm.
Errors are indicated in RRM in the Diagnostics window.
Table 7-5. Error messages.
Message
Description
Action
RAM error
An error in the gauge data memory
(RAM) has been detected during the
startup tests. Note: this automatically
resets the gauge.
Contact Emerson Process
Management Service Department.
FPROM error
An error in the gauge program
memory (FPROM) has been detected
during the startup tests. Note: this
automatically resets the gauge.
Contact Emerson Process
Management Service Department.
Database (Hreg) error
An error in the transmitter
configuration memory (EEPROM) has
been detected. The error is either a
checksum error that can be solved by
loading the default database or a
hardware error.
NOTE: the default values are used
until the problem is solved.
Load default database and restart the
transmitter. Contact Emerson
Process Management Service
Department if the problem persists.
Microwave Module error
An error in the microwave module.
Contact Emerson Process
Management Service Department.
LCD error
An error in the LCD is detected.
Contact Emerson Process
Management Service Department.
Modem error
An error in the modem used for digital
communication has been detected.
Contact Emerson Process
Management Service Department.
Analog out error
An error in the Analog Out Module.
Contact Emerson Process
Management Service Department.
Internal temperature error
An error in the internal temperature Contact Emerson Process Other hardware error An unspecified hardware error has Contact Emerson Process Measurement error A serious measurement error has Contact Emerson Process Configuration error At least one configuration parameter • Load the default database and restart the transmitter (see “Reset • Configure the transmitter or upload a backup configuration file (see • Contact Emerson Process Management Service Department if Software error An error has been detected in the Contact Emerson Process
measurement.
-40 ºC
Management Service Department.
been detected.
Management Service Department.
been detected.
Management Service Department.
is outside allowed range.
NOTE: the default values are used
until the problem is solved.
to Factory Settings” on page 7-22)
“Backing up the Transmitter
Configuration” on page 7-20)
the problem persists
transmitter software.
Management Service Department.