Reference Manual
00809-0100-4530, Rev BA
July 2009
Rosemount 5300 Series
D-4
REMOTE HOUSING
CONFIGURATION
When using remote housing, the remote connection length should be
configured. If the Remote Housing is ordered with a transmitter, it is
configured in the factory.
In Rosemount Radar Master, select Setup > Tank.
Under the Probe tab, Remote Housing length can be selected. Click the Store
button to save changes.
In the 375 Field Communicator, Remote Housing can be set with HART
command [2, 3, 1, 4, 2].
NOTE!
When disconnecting a remote housing cable from a probe in low
temperatures, -22 °F (-30 °C), Radar Master may not show that the probe is
missing.