Channel Definitions

Each input has an assigned channel which can be linked to the AI block. The
channels for the Rosemount 5300 Series are the following:

Table F-1. Channel
Assignments

PARAMETERS AND
DESCRIPTIONS

Table F-2. Level Transducer
Block Parameters and
Descriptions.

Channel Name

Channel

Number

Process variable

Level

1

CHANNEL_RADAR_LEVEL

Ullage

2

CHANNEL_RADAR_ULLAGE

Level Rate

3

CHANNEL_RADAR_LEVELRATE

Signal Strength

4

CHANNEL_RADAR_SIGNAL_STRENGTH

Volume

5

CHANNEL_RADAR_VOLUME

Internal Temperature

6

CHANNEL_RADAR_INTERNAL_TEMPERATURE

Upper Product Volume

7

CHANNEL_UPPER_PRODUCT_VOLUME

Lower Product Volume

8

CHANNEL_LOWER_ PRODUCT_VOLUME

Interface Distance

9

CHANNEL_INTERFACE_ DISTANCE

Upper Product Thickness

10

CHANNEL_UPPER_ PRODUCT_THICKNESS

Interface Level

11

CHANNEL_INTERFACE_LEVEL

Interface Level Rate

12

CHANNEL_INTERFACE_ LEVELRATE

Interface Signal Strength

13

CHANNEL_INTERFACE_ SIGNALSTRENGTH

Signal Quality

14

CHANNEL_SIGNAL_QUALITY

Surface/Noise Margin

15

CHANNEL_ SURFACE_NOISE_MARGIN

Vapor DC

16

CHANNEL_VAPOR_DC

Parameter

Index Number

Description

ST_REV

1

The revision level of the static data associated with the
function block. The revision value increments each time a
static parameter value in the block is changed.

TAG_DESC

2

The user description of the intended application of the block.

STRATEGY

3

The strategy field can be used to identify grouping of blocks.
This data is not checked or processed by the block.

ALERT_KEY

4

The identification number of the plant unit. This information
may be used in the host for sorting alarms, etc.

MODE_BLK

5

The actual, target, permitted, and normal modes of the block.
Target: The mode to “go to”
Actual: The mode the “block is currently in”
Permitted: Allowed modes that target may take on
Normal: Most common mode for target

BLOCK_ERR

6

This parameter reflects the error status associated with the
hardware or software components associated with a block. It
is a bit string, so that multiple errors may be shown.

UPDATE_EVT

7