Rosemount 5300 series – Emerson Process Management ROSEMOUNT 5300 User Manual
Page 234
Reference Manual
00809-0100-4530, Rev BA
July 2009
Rosemount 5300 Series
B-12
Figure B-10. Approval Labels
Factory Mutual (FM) Foundation
Fieldbus model
I5
Intrinsically Safe for Class I, II, III, Division 1, Groups A, B, C, D, E, F
and G.
Class I, Zone 0, AEX ia IIC T4 when installed per Control Drawing:
9240030-936.
Non-incendive for Class I, Division 2, Groups A, B, C and D.
Suitable for Class II, III, Division 2, Groups F and G;
F
OUNDATION
Fieldbus model: U
i
=30 Vdc, I
i
=300 mA, P
i
=1.3 W, C
i
=0 nF,
L
i
=0 H. Max operation 32 V, 25 mA.
Temperature code T4.
Ambient Temperature Limits: -50
°
C to + 60
°
C
Figure B-11. Approval Labels
Factory Mutual (FM) FISCO
model
IE
Intrinsically Safe for Class I, II, III, Division 1, Groups A, B, C, D, E, F
and G.
Class I, Zone 0, AEX ia IIC T4 when installed per Control Drawing:
9240030-936.
FISCO model: U
i
=17.5 Vdc, I
i
=380 mA, P
i
=5.32 W, L
i
=C
i
=0.
Temperature code T4.
Ambient Temperature Limits: -50
°
C to + 60
°
C