background image

Reference Manual

00809-0100-4530, Rev BA

July 2009

Rosemount 5300 Series

B-12

Figure B-10. Approval Labels
Factory Mutual (FM) Foundation
Fieldbus model

I5

Intrinsically Safe for Class I, II, III, Division 1, Groups A, B, C, D, E, F
and G.
Class I, Zone 0, AEX ia IIC T4 when installed per Control Drawing:
9240030-936.

Non-incendive for Class I, Division 2, Groups A, B, C and D.

Suitable for Class II, III, Division 2, Groups F and G;

F

OUNDATION

Fieldbus model: U

i

=30 Vdc, I

i

=300 mA, P

i

=1.3 W, C

i

=0 nF,

L

i

=0 H. Max operation 32 V, 25 mA.

Temperature code T4.
Ambient Temperature Limits: -50

°

C to + 60

°

C

Figure B-11. Approval Labels
Factory Mutual (FM) FISCO
model

IE

Intrinsically Safe for Class I, II, III, Division 1, Groups A, B, C, D, E, F
and G.
Class I, Zone 0, AEX ia IIC T4 when installed per Control Drawing:
9240030-936.

FISCO model: U

i

=17.5 Vdc, I

i

=380 mA, P

i

=5.32 W, L

i

=C

i

=0.

Temperature code T4.
Ambient Temperature Limits: -50

°

C to + 60

°

C