Rosemount 5300 series, Reference manual, Flange connection rating – Emerson Process Management ROSEMOUNT 5300 User Manual
Page 190: Flange rating standards
Reference Manual
00809-0100-4530, Rev BA
July 2009
Rosemount 5300 Series
A-6
Table A-1. Temperature range
for different tank seal material.
For Tri-Clamps, the maximum pressure is 16 bar for 1.5 in. (37.5 mm) and
2 in. (50 mm) housing; and 10 bar for 3 in. (75 mm) and 4 in. (100 mm)
housing. The final rating depends on the clamp and gasket you are using.
Tri-Clamp is available for the Standard Temperature and Pressure seal.
Flange Connection
Rating
Flange strength calculations are made with the following conditions:
•
For 316L SST:
•
For Alloy C-276:
Calculations show that the following applies.
Flange Rating Standards
•
ANSI:
316L SST Flanges according to ANSI B16.5 Table 2-2.3.
Standard: Max. 302 °F/580 psig
(150 °C/40 Bar).
HP/HTHP/C: Up to Class 2500.
Alloy C-276 (UNS N10276) flanges
according to ANSI B16.5 Table 2-3.8.
HP: Class 1500 up to max 200 °C or
max 100 °F/5000 psig (38°C/345 bar) and
392 °F/3500 psig (200 °C/243 bar)
HTHP: Class 1500 up to max 400 °C or max
100 °F/5000 psig (38 °C/345 bar) and
752 °F/2940 psig (400 °C/203 bar).
Tank Seal with different
O-ring material
Min. Temperature
°F (°C) in air
Max. Temperature
°F (°C) in air
Viton
5 (-15)
302 (150)
Ethylene Propylene (EPDM)
-40 (-40)
266 (130)
Kalrez 6375
14 (-10)
302 (150)
Buna-N
-31 (-35)
230 (110)
Bolting material
Gasket
Flange material
Hub material
Standard/HTHP
HP/HTHP/C
ANSI
Stainless steel SA193
B8M C1.2
Soft (1a) with min.
thickness 1.6 mm.
Spiral wound gasket with
nonmetallic filler (1b)
Stainless steel A182
Gr. F316L and
EN 10222-5-1.4404.
Stainless steel A479M
316L or
EN 10272-1.4404.
EN
EN 1515-1/-2 group
13E0, A4-70.
Soft (EN 1514-1) with
min. thickness 1.6 mm.
Spiral wound gasket with
nonmetallic filler
(EN 1514-2)
Bolting material
Gasket
Flange material
Hub material
HTHP
HP/HTHP/C
ANSI
UNS N10276
Soft (1a) with min.
Thickness 1.6 mm
Spiral wound gasket with
nonmetallic filler (1b)
SB462 Gr. N10276
(solution annealed
condition) or SB575 Gr.
N10276 (solution
annealed condition)
SB574 Gr. N10276
EN
Soft (EN 1514-1) with
min. Thickness 1.6 mm
Spiral wound gasket with
nonmetallic filler (EN
1514-2)