Rosemount 5300 series – Emerson Process Management ROSEMOUNT 5300 User Manual
Page 232
Reference Manual
00809-0100-4530, Rev BA
July 2009
Rosemount 5300 Series
B-10
Figure B-8. Approval Label
National Supervision and
Inspection Center for Explosion
Protection and Safety of
Instrumentation (NEPSI)
FISCO Model
IC
The following information is provided as part of the label of the
transmitter:
•
Ex ia IIC T4 (- 50 °C < T
a
< + 60 °C)
•
U
i
= 17.5 V, I
i
= 380 mA, P
i
= 5.32 W, C
i
0 nF, L
i
0
Special Conditions for Safe Use (X):
Refer to Certificates: GYJ081080X for Ex ia IIC T4 and GYJ081130X for
Ex iad IIC T4.