Reference Manual

00809-0100-4530, Rev BA

July 2009

Rosemount 5300 Series

D-2

M6 screw

Fasten the housing support to the bracket

Fasten the housing support to the bracket
using the M6 screws. The screws are
threaded through the top of the mounting
bracket and into the housing support.

M50 nut

Mount the probe housing on the probe

Mount the probe housing on the probe,
making sure that the M50 nut is properly
tightened.

Connect the transmitter head on the
housing support

Connect the transmitter head on the housing
support, making sure that the M50 nut is
properly tightened.

See “Remote Housing Configuration” on
page D-4 to check
if the transmitter is
configured with remote housing.

Mount the bracket to the pole

Mount the bracket to the pole, making sure the
distance between the probe and bracket does
not exceed the length of the remote connection.

1. Put the two U-bolts through the holes of

the bracket. Several holes are available for
vertical/horizontal pipe mounting.

2. Put the clamping brackets on the U-bolts

and around the pipe.

3. Use the supplied nuts to fasten the bracket

to the pipe.

U-bolt

Bracket