Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual
Page 93

•
In the Device Status area, if the icon changes to
and the message
remains, the timer counts down a fixed pause interval of 60 seconds. The alarm
tone returns after the pause interval.
A longer pause interval can be configured in the Advanced tab. You start the
configured pause interval by touching
.
If you responded to an NIBP alarm and multiple NIBP limits have been
exceeded, the alarm tone pauses, the first message clears, and the next NIBP
limit message appears with a countdown timer. A new NIBP alarm tone sounds
after the countdown unless you touch
to dismiss each remaining NIBP
limit message.
2. If multiple alarms are active, a multiple alarm toggle appears in the Device Status
area. The monitor will either autoscroll through the alarm messages, displaying each
for approximately 4 seconds in order of priority, or you can move through the alarms
manually. Respond to multiple alarms manually as follows:
a. Touch
to pause all audio alarms.
The alarm icon changes to
and the timer counts down a fixed pause
interval of 60 seconds. The alarm tone returns after the pause interval.
b. Touch
in the Device Status area to view each alarm in the stack.
Note
The multiple alarm toggle button displays the number of
active alarms inside the alarm icon. A set of dots
indicating the display order of alarms from highest (left)
to lowest (right) priority (as well as the most recent in
the case of multiple alarms of the same priority) appears
below it.
Directions for use
Alarms 87