beautypg.com

Set response mode, Measure spo2 and pulse rate – Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual

Page 152

background image

3. Touch the SpO2 vertical tab.
4. Ensure that the SpO2 alarm limit control is set to ON.

Note

If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals will occur for that specific parameter.

5. Enter the desired upper and lower alarms limits for SpO2 using the up/down arrow

keys or the keypad.

6. If your monitor is configured with a Nellcor SpO2 sensor, touch

to

select a SatSeconds setting.

7. Touch the Home tab.

The new alarm settings take effect immediately.

Set Response Mode

To set the Response Mode from the Home tab, the monitor must be in the Intervals

Monitoring or Continuous Monitoring profile.

Touch

in the SpO2 frame.

When Fast mode is selected, either MODE: Fast or : Fast appears in the frame.

Measure SpO2 and pulse rate

WARNING Inaccurate measurement risk. Use only Masimo rainbow SET
sensors and accessories on Masimo-equipped monitors.

WARNING Inaccurate measurement risk. Use only Nellcor sensors and
accessories on Nellcor-equipped monitors.

WARNING High-intensity lights (including pulsating strobe lights) directed
on the pulse co-oximeter can prevent it from obtaining readings.

WARNING The pulsations from intra-aortic balloon support can increase
the pulse rate displayed on the monitor. Verify the patient's pulse rate
against the ECG heart rate.

WARNING Inaccurate measurement risk. Venous congestion might cause
a low reading of arterial oxygen saturation. Place the sensor on a hand at
heart level to ensure proper venous outflow from a monitored site.

WARNING Inaccurate measurement risk. The pulse co-oximeter can be
used during defibrillation, but the reading may be inaccurate for up to 20
seconds.

WARNING Patient injury risk. Do not attempt to reprocess, recondition, or
modify any sensors or patient cables. Doing so might damage electrical
components.

146 Patient monitoring

Welch Allyn Connex

®