Specify server settings – Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual
Page 184

Note None of the changed radio settings take effect
until you touch
Configure radio
.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify server settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Network tab.
3. Touch the Server tab.
4. Select the method used to identify the IP address of the server with which the
device will communicate.
5. Specify settings.
Setting
Action/Description
Broadcast
Enable the device to broadcast a request to obtain an IP
address for a selected service. Specify the port to match
the port used by the server. Touch
in the Port entry
field and enter the port number. The range of entry is 0
to 65535.
VM IP
Enable the device to connect to a Connex Vitals
Management (VM) Datacatcher or other server at a fixed
IP address. Touch
in the Vitals Management server
IP address fields and enter the IP address. Touch
in
the Port entry field and enter the port number. The range
of entry is 0 to 65535.
NRS IP
Enable the device to connect to a Network Rendezvous
Service (NRS) at a fixed IP address. Touch
in the
Network rendezvous service IP address fields and enter
the IP address. Touch
in the Port entry field and
enter the port number. The range of entry is 0 to 65535.
DNS name
Enable the device to connect to a Network Rendezvous
Service (NRS) by entering a hostname to be sent to a
Domain Name Server (DNS) to retrieve the NRS IP
address. Touch
in the Network rendezvous service
DNS name field and enter the DNS name. Touch
in
the Port entry field and enter the port number. The range
of entry is 0 to 65535.
Note The DNS name option is only available in the
following conditions:
178 Advanced settings