Continuous patient monitoring

Your device must be configured with the Continuous Monitoring profile to perform

continuous patient monitoring.

Key terms

Patient context

Condition in which a patient ID and patient type have been selected on a device.

Location context

Condition in which a room or a room and bed have been selected on a device.

You have the option to continuously monitor patients on the device alone (primary

monitoring) or to stream continuous patient data to a central station as well (secondary

monitoring).

If Continuous Monitoring has been configured as the default profile, then the device

starts up in this profile. If the Continuous Monitoring profile is active, you can proceed

with an existing patient and/or location context, or you can assign a patient and location.

Activate the Continuous Monitoring profile

If the Continuous Monitoring profile is not active, follow these steps:

1. Touch the Settings tab.
2. Touch the Device tab.

The Profile vertical tab appears.

3. Select Continuous Monitoring.

When neither patient nor location context are established and no patient

measurements have been taken or saved, the Continuous Monitoring profile

immediately becomes active. If these conditions do not apply, proceed to step 4.

4. Using the conditions presented in boldface text, select and follow the steps below

that apply to your patient and the device.
a. When patient and/or location context are established but no patient

measurements have been taken or saved:
A confirmation dialog appears asking you to confirm the patient and/or location

context information or select a new patient. For example, if both patient and

location context have been established, the following options appear:

Same patient, same location

Same patient, different location

New patient

Select the desired option and touch OK.

b. When patient and/or location context are established and patient

measurements have been taken or saved:
A confirmation dialog appears with the following message: "Changing to the

Continuous Monitoring profile deletes all measurement data from the device.

Continue anyway?"

1. Touch Yes to confirm or No to cancel the profile change.

54 Connex CS