USB and USB flash drive messages

Message

Possible cause

Suggested action

USB Communication failure. Call

for service.

An internal or external device is connected but

failed enumeration

Power down and restart.

Check USB connections.

If problem persists, call for

service.

External device not licensed for

use.

A license for an external device (e.g., barcode

scanner) has not been activated

Disconnect the unlicensed

device.

Call for service to obtain the

authorization code from Welch

Allyn to activate the license.

External device not recognized.

An unrecognized external device is connected

Disconnect the unrecognized

device.

Incompatible Welch Allyn device. A communication protocol failure has occurred

Call for service.

USB accessory disconnected.

The USB cable between an external device and

the monitor is disconnected

Confirm that the USB cable is

connected to the device and the

monitor.

Save not successful.

Missing, improperly inserted, or incompatible

USB flash drive

Dismiss message and insert a

compatible USB flash drive.

Unable to save configuration to

USB.

Missing, improperly inserted, or incompatible

USB flash drive

Dismiss message and insert a

compatible USB flash drive.

System messages

Message

Possible cause

Suggested action

Set date and time.

The date or time is not set

Set the date and time.

The date or time is not set properly

Reset the date or time.

Device shutdown is not available

at this time.

Device cannot perform an immediate shutdown Touch

OK

, wait, and retry.

Advanced settings unavailable.

Sensors are taking measurements

Stop continuous measurements.

A physiological alarm condition is active

Respond to or reset the alarm.

Spot Check measurements have not been saved Save the measurements.

Unexpected restart occurred. Call

for service.

A system error caused the monitor to restart.

Call for service.

190 Troubleshooting