CO2 specifications (for additional information, refer to the directions for use that accompanies
each accessory)
Sampled gas return/disposal
Sampled gases are not returned to the breathing circuit.
Exhaled gas is exhausted from the monitor's exhaust port.
Dispose of sampled gases according to facility requirements
or local regulations.
Calibration interval
Initial: After 1200 operating hours
Subsequent: After 4000 operating hours or annually
(whichever comes first)
Periodic service
After 30,000 operating hours
etCO2
Units of measure
mmHg, kPa; user-selectable
Display range
0 to 150 mmHg (0.0 to 20.0 kPa)
Resolution
1 mmHg, 0.1 kPa
FiCO2
Units of measure
mmHg, kPa; user-selectable
Display range
0 to 150 mmHg (0.0 to 20.0 kPa)
Resolution
1 mmHg, 0.1 kPa
IPI
Display range
1 to 10
RR
Unit of measure
breaths per minute
Display range
0 to 150 breaths per minute
Accuracy
0 to 70 breaths per minute: ±1 breath per minute
71 to 120 breaths per minute: ±2 breaths per minute
121 to 150 breaths per minute: ±3 breaths per minute
Resolution
1 breath per minute
1
The accuracy specification is maintained to within 4% of the values indicated in the presence of interfering gases in
accordance with ISO 21647, clauses 51.101.3 and 101.1.
2
The accuracy specification is maintained to within 4% of the values indicated in the presence of up to 80% helium
with up to 15% oxygen when tested in accordance with ISO 21647.
3
The accuracy specification is maintained when tested for drift in accordance with ISO 21647, clause 51.101.2.
Directions for use
Specifications 203