Specify clinician settings
Clinician identification appears next to the medicine symbol in the Device Status area on
the Home tab.
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Data Management tab.
3. Touch the Clinician tab.
4. Specify settings.
Setting
Action/Description
Label
Select a type of clinician identification label for display
on the Home tab:
Full name
,
Abbreviation
,
Clinician ID
, or
Symbol only
.
Require clinician ID to save readings
Make entering a clinician ID a prerequisite for saving
measurements. If they fail to enter identification, the
monitor prompts them when they try to save
measurements. Clinicians can enter clinician
identification on the Clinician tab.
Clear clinician information on manual save
Specify that the monitor clears the selected clinician
after a clinician manually saves measurements from the
Home tab. Clinician information clears from the Clinician
tab and the Device Status area.
Search by clinician ID
Enable the monitor to query the network for clinician
information based on ID. The monitor initiates the search
when the clinician enters or scans the ID from the
Clinician tab. Returned clinician information populates
the Device Status area and fields on the Clinician tab.
Select
Require password
to require clinicians to
enter their password, in addition to ID, on the Clinician
tab. The monitor uses the ID and password combination
to query the network for clinician information.
Select
Require clinician ID match to save
measurements
to require that the selected
clinican ID matches a clinician ID in an external host
system before you can save measurements.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch Exit.