Note

To switch to the Direct mode, touch

after the

Predictive mode measurement is acquired. The temperature
frame displays text indicating that the Direct mode is active.
The monitor sounds a tone to indicate the start of a Direct
measurement. Once you are in the Direct mode, continue to
separate the buttocks and hold the probe in place throughout
the measurement process.

Note

The monitor does not retain Direct mode temperatures in
memory. Therefore, it is important to note the temperature
before removing the probe from the measurement site and
then manually record it in the patient record.

6. Remove the probe after the temperature measurement is complete and firmly press

the eject button on the top of the probe to release the probe cover.

7. Return the probe to the probe well.
8. Wash your hands to reduce the risk of cross-contamination.

Braun ThermoScan

®

PRO 4000 thermometer and accessory dock

The thermometer and accessory dock enable you to transfer an ear temperature

measurement to the monitor. The dock also charges the thermometer battery.

Read the thermometer manufacturer's directions for use before attempting to configure,

use, troubleshoot, or maintain the thermometer.

WARNING Liquids can damage electronics inside the thermometer.
Prevent liquids from spilling on the thermometer. If liquids are spilled on
the thermometer, dry off the thermometer with a clean cloth. Check for
proper operation and accuracy. If liquids possibly entered the thermometer,
remove the thermometer from use until it has been properly dried,
inspected, and tested by qualified service personnel.

CAUTION Probe covers are disposable, nonsterilized, and single-use. The
thermometer is also nonsterilized. Do not autoclave the thermometer and
probe covers. Ensure that probe covers are disposed of according to facility
requirements or local regulations.

CAUTION The thermometer has no user-serviceable parts. If service is
required, call your nearest Welch Allyn Customer Service or Technical
Support facility.

CAUTION Store the thermometer and probe covers in a dry location, free
from dust and contamination and away from direct sunlight. Keep the
ambient temperature at the storage location fairly constant and within the
range of 50°F to 104°F (10°C to 40°C).

Take a temperature at the ear site

WARNING Probe covers are single-use only. Re-use of a probe cover may
result in spread of bacteria and cross-contamination.

140 Patient monitoring

Welch Allyn Connex

®