Page 76

background image

Same patient, different location

New patient

b. Select the desired option and touch OK.

The profile change immediately takes effect.

4. When patient and/or location context are established and patient

measurements have been taken or saved:
a. Select the desired profile.

A confirmation dialog appears with the following message: "Changing to the

Continuous Monitoring profile deletes all measurement data from the device.

Continue anyway?"

b. Touch Yes to confirm or No to cancel the profile change.

Note

Selecting No enables you to send captured vitals data on
the device before changing profiles. After you complete
that task, if applicable, repeat steps 1 and 4.

A confirmation dialog appears asking you to confirm the patient and/or location

context information or select a new patient. For example, if both patient and

location context have been established, the following options appear:

Same patient, same location

Same patient, different location

New patient

c. Select the desired option and touch OK.

The profile change immediately takes effect.

5. Touch the Home tab.

You can now start monitoring a patient in the Continuous Monitoring profile.

Change from the Continuous Monitoring profile

To change from the Continuous Monitoring profile to any other profile, complete these

steps.

1. Touch the selected profile indicator in the Device Status area.

The Profile vertical tab appears. Based on the boldface text presented, select and

follow the step and substeps below that apply to your patient and the device.

2. When no patient trend data has been stored, select the desired profile.

The profile change immediately takes effect.

3. When continuous sensors are connected to the patient, or when episodic or

continuous data has been stored, no other profile selections are available until you

remove patient sensors and end monitoring.
a. Remove continuous sensors from your patient.
b. On the Settings tab, touch the Monitor tab.
c. Touch End monitoring.

An "End monitoring" dialog appears with the following options:

70 Profiles

Welch Allyn Connex

®

Vital Signs Monitor 6000 Series™