Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual
Page 134

Note
In the Continuous Monitoring profile, all patient measurements remain
onscreen when automatic interval measurements are saved.
The
button changes to a timer (
), which counts down to the next
automatic measurement.
Automatic measurements continue until you turn off intervals.
WARNING Patient harm risk. Do not use intervals on neonates out of
earshot. Verify that audio can be heard from where you intend to be.
Set up NIBP intervals
Follow these steps to configure NIBP intervals.
1. Touch the Settings tab.
2. Touch the Setup tab.
3. Touch the Intervals vertical tab.
4. Select Automatic, Program, or Stat, and follow the steps presented in the
associated topics that follow.
5. To automatically print patient data at each interval, touch the Automatic print on
interval checkbox.
6. To start intervals immediately, touch Start intervals. Otherwise, touch the Home
tab.
The new settings take effect immediately.
Automatic intervals
You can configure the monitor to take automatic NIBP measurements at consistent
intervals. Use the spin box or keypad to set the desired interval.
Note
An alarm does not turn off intervals. Subsequent automatic measurements
continue to occur as scheduled.
128 Patient monitoring
Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™