Network messages, Problems and solutions – Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual

Page 198

background image

Message

Possible cause

Suggested action

Printer not functional. Call for

service.

The printer motor is broken

Call for service.

The detection switch malfunctioned

A hardware failure occurred in the power supply

The printer does not identify itself correctly

The printer does not enumerate

Printing records:

Note The number of records

requested appears in the message

and counts down during printing.

The monitor is printing the records selected on

the Review tab

Acknowledge the number of

records printing or touch Cancel

to interrupt printing.

Printing report; please wait.

The printer needs more time to complete a print

job when the Automatic print on interval

control has been enabled

Wait for the print job to print

completely.

Disable Automatic print on

interval in Intervals settings.

Network messages

Message

Possible cause

Suggested action

Lost connectivity with host.

The central station or other host

application has stopped running or

is not operating properly

Confirm that the host application

is running properly.

Problems and solutions

The problems addressed in this table do not generate alarm or information messages on

the monitor.

Problem

Possible cause

Suggested action

No SpHb value is displayed

An SpO2-only cable is connected to the monitor Replace the SpO2-only cable

with an SpO2/SpHb (Masimo

rainbow) cable.

The SpO2/SpHb reusable sensor has expired

Note

A technical alarm appears.

Replace the sensor.

Poor sensor placement on the patient

Remove the sensor from the

patient and reapply.

The monitor may have the SpHb license, but the

SpO2 module does not

Contact Welch Allyn to verify

that the SpO2 module contains

the SpHb license.

192 Troubleshooting

Welch Allyn Connex

®

Vital Signs Monitor 6000 Series™