Network messages, Problems and solutions – Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual
Page 198

Message
Possible cause
Suggested action
Printer not functional. Call for
service.
The printer motor is broken
Call for service.
The detection switch malfunctioned
A hardware failure occurred in the power supply
The printer does not identify itself correctly
The printer does not enumerate
Printing records:
Note The number of records
requested appears in the message
and counts down during printing.
The monitor is printing the records selected on
the Review tab
Acknowledge the number of
records printing or touch Cancel
to interrupt printing.
Printing report; please wait.
The printer needs more time to complete a print
job when the Automatic print on interval
control has been enabled
Wait for the print job to print
completely.
Disable Automatic print on
interval in Intervals settings.
Network messages
Message
Possible cause
Suggested action
Lost connectivity with host.
The central station or other host
application has stopped running or
is not operating properly
Confirm that the host application
is running properly.
Problems and solutions
The problems addressed in this table do not generate alarm or information messages on
the monitor.
Problem
Possible cause
Suggested action
No SpHb value is displayed
An SpO2-only cable is connected to the monitor Replace the SpO2-only cable
with an SpO2/SpHb (Masimo
rainbow) cable.
The SpO2/SpHb reusable sensor has expired
Note
A technical alarm appears.
Replace the sensor.
Poor sensor placement on the patient
Remove the sensor from the
patient and reapply.
The monitor may have the SpHb license, but the
SpO2 module does not
Contact Welch Allyn to verify
that the SpO2 module contains
the SpHb license.
192 Troubleshooting
Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™