Note
If no patient is selected, the Continuous Monitoring profile is not active,
and no measurements have been taken, all table cells in the Review tab
remain blank.
Note
In the Continuous Monitoring profile, you cannot delete saved patient
measurements. Patient measurements older than 24 hours are
automatically deleted from the Review tab.
Note
The date and time stamps on saved patient measurements will adjust in
response to new date and time settings.
Note
In the Continuous Monitoring profile, the Send option is not available on the
Review tab.
Manage patient records (Intervals Monitoring, Spot
Check, and Triage profiles)
Patient records can be sent to the network, printed, or deleted in these profiles.
Note
Because the Patient tab is not available in the Triage profile, records saved
in the Triage profile contain only vital signs measurements. Even without
patient identification, these records can be sent to the network, printed, or
deleted.
1. Touch the Review tab.
Note
Measurements that triggered a physiological alarm are
highlighted on this tab.
2. Select records by touching the check box.
3. Touch Send to transmit the records to the network, Print to print the records, or
Delete to permanently remove the records as desired.
CAUTION Verify patient identity on the monitor after
manual or barcode entry and before printing or transferring
patient records.
CAUTION Always visually verify the printed patient records.