CAUTION Weight scales attached to this monitor must be running on
battery power (battery type is specified in the weight scale manufacturer's
directions for use). Do not use the weight scale's external power supply.
1. From the Home tab, touch anywhere within the Manual parameters frame.
The Manual tab appears.
2. Touch the keypad icon in selected fields to open the numeric keypad, and then
manually adjust height, weight, pain level, temperature, respiration rate, or other
parameters.
Note
If an approved, battery-powered weight scale is attached to
the monitor, measurements from the weight scale populate
fields in the Manual tab. You can adjust weight and height
measurements on this tab, but if you do, the read-only BMI
field will clear.
Note
Ensure that the current patient ID is correct before saving.
3. Touch OK to confirm settings and return to the Home tab.
Note
In the Intervals Monitoring and Spot Check (episodic)
profiles, manual parameters clear from the Home tab after a
successful save.
Note
When the device is connected to a central station and in the
Continuous Monitoring profile, all manual parameters except
temperature will be sent to the central station when you
touch OK.
4. Touch OK to confirm settings and return to the Home tab.
Note
In the Intervals Monitoring and Spot Check (episodic)
profiles, manual parameters clear from the Home tab after a
successful save.
Directions for use
Patient monitoring 155