Specify advanced nibp settings – Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual
Page 177

3. Touch the SpHb tab.
4. Specify settings.
Setting
Action/Description
Default view
Select a numeric view or a trend graph view as the
primary SpHb display on the Home tab.
Unit of measure
Select the primary unit of measure for the SpHb display
on the Home tab.
Default averaging
Select the default moving window of time used by the
parameter to calculate the SpHb value and update the
display: short (approximately 1 minute), medium
(approximately 3 minutes), or long (approximately 6
minutes).
Reference
Select arterial or venous as the calibrated reference
source.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced tabs and return to the Home tab, touch Exit.
Specify advanced NIBP settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Parameters tab.
3. Touch the NIBP tab.
4. Specify settings.
Setting
Action/Description
Default view
Select primary and secondary views.
Select
Display MAP
to display mean arterial
pressure (MAP) in the NIBP frame on the Home tab.
If
Display MAP
is selected, specify which numerics
are primary in the NIBP frame. On the Home tab,
clinicians can touch the NIBP frame to toggle between
views.
Unit of measure
Select the NIBP unit of measure for display.
Tube type
Select the number of tubes that are connected to the
NIBP cuff used with this monitor. If you select
1 tube
,
the only algorithm available for selection is
Step
.
Algorithm default
Select the default algorithm used to determine NIBP
measurements.
Cuff inflation target
If you select the
Step
algorithm, touch
and enter
a default cuff inflation target for each type of patient.
Clinicians can change the CITs from the default CITs that
Directions for use
Advanced settings 171