Message
Possible cause
Suggested action
Note This message only appears
on devices configured with SpHb.
Temperature messages
Message
Possible cause
Suggested action
Connect temperature probe.
No probe is connected
Connect a temperature probe and
retry.
The probe is faulty
Replace the temperature probe.
The temperature module returned a connect
probe message
Connect a temperature probe and
retry. If a probe is already
connected, replace the probe.
Insert correct color-coded probe
well.
The probe well is missing
Insert a temperature probe well.
Replace temperature probe.
The probe is faulty
Replace the temperature probe.
Temperature not functional. Call
for service.
A module error occurred
Call for service.
Temperature time limit exceeded.
Retry temperature measurement.
The direct mode timed out
Remove the probe from the
measurement site.
Tissue contact lost.
The probe has lost contact with the patient's
tissue
Touch OK to dismiss the
message. Return the probe to the
probe well and retry patient
temperature measurement.
Ensure that the probe makes
proper contact with the patient's
tissue.
Retry temperature measurement.
Note This message often
accompanies other temperature
messages.
A probe heater or data error occurred
Retry the temperature
measurement. If the problem
persists, replace the probe.
User settings require adjustment
Adjust the user settings and
retry.
The ambient temperature is out of range
Operate the monitor within the
specified temperature range.
Retry patient temperature
measurement.
The SureTemp temperature module is faulty
Call for service.
Unable to detect new temperature.
Retry measurement.
The Braun thermometer is removed from the dock
or not seated properly in the dock
Return the thermometer to the
dock or adjust the thermometer
in the dock.
The Braun dock is disconnected from the device Connect the USB cable for the
Braun dock to the device.
Directions for use