Temperature messages

Message

Possible cause

Suggested action

Connect temperature probe.

No probe is connected

Connect a temperature probe and

retry.

The probe is faulty

Replace the temperature probe.

The temperature module returned a connect

probe message

Connect a temperature probe and

retry. If a probe is already

connected, replace the probe.

Insert correct color-coded probe

well.

The probe well is missing

Insert a temperature probe well.

Replace temperature probe.

The probe is faulty

Replace the temperature probe.

Temperature not functional. Call

for service.

A module error occurred

Call for service.

Temperature time limit exceeded.

Retry temperature measurement.

The direct mode timed out

Remove the probe from the

measurement site.

Tissue contact lost.

The probe has lost contact with the patient's

tissue

Touch OK to dismiss the

message. Return the probe to the

probe well and retry patient

temperature measurement.

Ensure that the probe makes

proper contact with the patient's

tissue.

Retry temperature measurement.

Note This message often

accompanies other temperature

messages.

A probe heater or data error occurred

Retry the temperature

measurement. If the problem

persists, replace the probe.

User settings require adjustment

Adjust the user settings and

retry.

The ambient temperature is out of range

Operate the monitor within the

specified temperature range.

Retry patient temperature

measurement.

The SureTemp temperature module is faulty

Call for service.

Unable to detect new temperature.

Retry measurement.

The Braun thermometer is removed from the dock

or not seated properly in the dock

Return the thermometer to the

dock or adjust the thermometer

in the dock.

The Braun dock is disconnected from the device Connect the USB cable for the

Braun dock to the device.

Directions for use

Troubleshooting 187