Take a temperature in the direct mode – Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual

Page 143

background image

While the measurement is taking place, the temperature frame displays the process

indicator.

5. The monitor sounds a tone when the final temperature is reached (in approximately

6 to 15 seconds). The temperature frame continues to display the temperature even

after the probe is returned to the probe well.

Note

To switch to the Direct mode, touch

after you

acquire the Predictive mode measurement. The temperature
frame displays text indicating that the Direct mode is active.

The monitor sounds a tone at the start of a Direct mode measurement.

6. Remove the probe after the temperature measurement is complete and firmly press

the eject button on the top of the probe to release the probe cover.

WARNING Patient injury risk. Do not exceed the
recommended temperature measurement durations in Direct
mode. Continuous measurement durations of 3 minutes at
the oral and rectal sites and 5 minutes at the axillary site are
recommended for accurate measurement. Do not
continuously measure beyond 10 minutes in any mode.

Ensure that probe covers are disposed of according to facility requirements or local

regulations.

7. Return the probe to the probe well.
8. Wash your hands to reduce the risk of cross-contamination.

Take a temperature in the Direct mode

Direct mode displays the temperature of the probe as long as the probe tip remains in

place at the measurement site and remains within the operating patient temperature

range. The patient’s temperature will reach final equilibrium in approximately 3 minutes

at the oral and rectal measurement sites and approximately 5 minutes at the axillary site.

The monitor enters Direct mode by the following methods.

After you complete a Predictive mode measurement, touch

to switch from

Predictive to Direct mode. The temperature frame displays text indicating that the

Direct mode is active.

Remove the probe from the probe well, load a probe cover, select a temperature

site, and expose the probe to ambient air for more than 60 seconds to switch the

monitor to Direct mode. The temperature frame displays text indicating that the

Direct mode is active.

Directions for use

Patient monitoring 137