Configure pulse rate alarms, Manual parameters frame, Enter manual parameters – Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual
Page 160

1. Touch the Settings tab.
2. Touch the Setup tab.
3. Touch the Pulse rate vertical tab.
4. Select the desired Tone volume.
5. Touch the Home tab.
The new settings take effect immediately.
Configure pulse rate alarms
Follow these steps to set alarm limits for pulse rate.
1. Verify that you are using the Intervals Monitoring or Continuous Monitoring profile.
2. Touch the Alarms tab.
3. Touch the Pulse rate vertical tab.
4. Ensure that the Pulse rate alarm limit control is set to ON.
Note
If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals will occur for that specific parameter.
5. Enter the desired upper and lower alarm limits for pulse rate using the up/down
arrow keys or the keypad.
6. Touch the Home tab.
The new alarm settings take effect immediately.
Manual parameters frame
The Manual parameters frame, located in the lower right of the Home tab, supports
manual entry of parameters and displays measurements taken by some accessories.
Note
Manual parameters are not available in the Triage profile.
Note
Body mass index (BMI) is only available with an attached weight scale that
calculates BMI.
Note
When a measurement is transferred from an attached weight scale to the
device, the measurement displayed on the device is within one decimal
place (0.1) of the measurement displayed by the weight scale.
Note
You cannot manually enter temperature in this frame on a device
configured with a SureTemp Plus temperature module.
Note
You cannot manually enter respiration rate in this frame on a device
configured with CO2, RRa, or EarlySense.
Enter manual parameters
Note
The Manual parameters frame enables you to enter measurements taken
manually and displays measurements taken by some accessories. You can
select and configure the parameters in Advanced settings. Only four
parameters appear in the Manual parameters frame.
154 Patient monitoring