Patient rest mode location on the monitor
Access the rest mode on the Alarms tab.
CAUTION Loss of power will cause the monitor to return to default
settings. Each time you power up the monitor, you must set alarm limits
appropriate for your patient.
Turn on patient rest mode
To turn on the patient rest mode from the device:
1. Touch the Alarms tab.
The Alarms screen appears.
2. On the Alarms screen, touch Patient rest mode on.
The patient rest mode is activated.
Turn off patient rest mode
To turn off the patient rest mode from the device:
1. Touch the Alarms tab.
The Alarms screen appears.
2. On the Alarms screen, touch either Alarm audio on or Alarm audio off.
The patient rest mode is deactivated.
Reset (pause or turn off) audio alarms
WARNING Do not pause or turn off an audible alarm if patient safety
might be compromised.
Audio alarm characteristics
•
After you reset an audio alarm, some tones do not return, but others return after a
pause interval if the condition that caused the alarm persists.
•
If a new alarm condition occurs during a pause interval, a new audio tone occurs.
•
If an audio alarm is not paused or turned off after a period of time, a buzzer
accompanies the tone.
Pause or turn off an audio alarm
1. In the Device Status area, touch
.
•
Visual indications remain in the parameter frame until the condition is corrected,
the next measurement is taken, or the alarm is cancelled.