b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Data Management tab.
3. Touch the Patient tab.
4. Specify settings.
Setting
Action/Description
Default patient type
Select a default patient type for this monitor. The patient
type shows in the Patient frame on the Home tab.
In the Summary tab on the Patients tab, clinicians can
change the displayed patient type from the default
patient type that you set here.
Name format
Select a format for all displayed patient names:
Full
name
or
Abbreviation
.
Primary label
Select the primary identification label for all displayed
patients.
Secondary label
Select a secondary identification label for patients. A
secondary label displays only on the Home tab, after the
primary label.
Require patient ID to save readings
Make entering a patient ID a prerequisite for saving
measurements. If the clinician fails to enter an identifier,
the monitor prompts the clinician when they try to save.
Search by patient ID
Enable clinicians to enter a patient ID to query for the
patient's information. If clinicians scan the ID onto the
Home tab or the Summary tab, the monitor queries the
patient list and the network. Returned patient
information populates the Patient frame on the Home tab
and fields on the Summary tab.
Select
Require patient ID match to save
measurements
to require that the selected
patient ID matches a patient ID in the Patient list of the
device or an external host system before you can save
measurements.
Clear patient information on manual save
Specify that the monitor clears the selected patient after
a clinician manually saves measurements from the Home
tab. Patient information clears from the Patient frame
and the Summary tab.
Note This setting does not take effect when intervals
are in progress.
Retrieve list
Enable the monitor to retrieve the patient list from the
network. When this option is selected, a Retrieve list
button replaces the Add button on the List tab.
Information from the network populates the List tab
when clinicians touch the Retrieve list button. Since
the Add button is not available, clinicians cannot add a
patient to the patient list.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch Exit.