Resume continuous monitoring, Pause time remaining, End continuous monitoring – Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual
Page 62

Resume continuous monitoring
You can resume continuous monitoring after a pause when patient sensors are
connected to a patient. The status of the pause timer determines your next step.
Pause time remaining
If the pause time limit has not expired (time remains on the countdown timer), resume
monitoring as follows:
1. Reconnect patient sensors to the patient as needed.
2. Touch Resume monitoring.
The Home tab appears and continuous monitoring resumes.
Pause time limit expired (patient sensors connected)
If the pause time limit has expired and you have already reconnected patient sensors to
the patient and/or the device, the Home tab appears and continuous monitoring
automatically resumes.
Pause time limit expired (patient sensors disconnected)
If the pause time limit runs out before you resume monitoring, the device exits Pause
mode, and alarms might occur.
Reconnect patient sensors to the patient and/or device and acknowledge alarm and
information messages as needed.
Continuous monitoring resumes.
End continuous monitoring
When you no longer need to monitor a patient continuously, follow these steps to end
monitoring:
1. Touch the Settings tab.
The Monitor tab appears.
2. Touch End monitoring.
An "End monitoring" dialog appears with the following options:
•
New patient – Clears patient and location context and review data after you
remove patient sensors
•
Power down – Clears patient and location context and review data, and powers
down the device
•
Cancel – Retains patient data and returns to Home tab
3. Select the desired option.
56 Connex CS
Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™