parameter. See "Configure movement alarms" to adjust alarm settings. Follow these

steps to set up movement modifiers, if applicable.

1. Touch the Patients tab.
2. Touch the Manual tab.
3. Scroll through the list to find the Movement section, then enter or select the

Movement modifier(s) as desired.

Note

If no modifiers are associated with this parameter, no
Movement section will appear.

4. Touch OK.

The Home tab appears. In the Continuous Monitoring profile, the modifiers are saved

with the next set of measurements sent to the network.

Configure movement alarms and patient turn reminder

Follow these steps to set alarm limits for patient movement.

1. Verify that you are using the Continuous Monitoring profile.

Note

If “Allow profile change” is enabled in Advanced settings,
the device attempts to shift automatically to the Continuous
Monitoring profile when you connect the sensor cable.

2. Touch the Alarms tab.
3. Touch the Movement vertical tab.
4. Ensure that the Exit alarm and Motion alarm controls are set to ON.

Note

If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals occur for that specific parameter.

5. Adjust the following settings as desired:

Exit sensitivity. Use the up/down arrow keys or the keypad to choose a

sensitivity level ranging from 1 to 6. Setting 1 is the least sensitive, and setting 6

is the most sensitive.

Note

Levels 5 and 6 are highly sensitive and might trigger an
alarm as a result of minimal movement while the patient
is in bed. Correct patient assessment is the key to
minimize false alerts.

Turn reminder. Select the desired turn schedule from the list box.

6. Touch the Home tab.

The new alarm settings take effect immediately.

Monitor patient movement

WARNING Patient safety risk. Do not use heating blankets. Heating
blankets could reduce the safety and efficacy of the system and void the
warranty.

102 Patient monitoring

Welch Allyn Connex

®