Message

Possible cause

Suggested action

The sensor is defective.

A sensor error occurred

Replace the sensor.

Trend change detected. Review

patient history.

Patient measurements have changed enough to

require attention

Check the patient and patient

history.

CO2 messages

Message

Possible cause

Suggested action

CO2 not functional. Call for service. A nonrecoverable communications error has

occurred

Call for service.

Filter line disconnected.

The sampling line is not connected to the monitor Connect a sampling line to the

monitor.

Check to ensure that the

sampling line plug is tightly

connected to the monitor.

Check for occlusion in gas line.

The purge cycle failed to clear a blockage in the

sampling line

Replace the sampling line.

CO2 temperature out of range. CO2

might not be accurate.

CO2 module temperature is out of range

Wait for the module temperature

to return to normal and the

module to reset before

calibrating CO2.

Move to a location where the

ambient temperature enables the

module temperature to return to

normal and the module to reset

before calibrating CO2.

Connect or clear filter line.

Calibration failed because sampling line is

blocked or kinked

Check sampling line for

obstruction or kinks. Replace the

filter line if necessary.

Module is not ready for calibration because the

sampling line is not connected to the monitor

Check to ensure that the

sampling line plug is tightly

connected to the monitor.

Purging filter line.

An occlusion was detected in the sampling line,

which automatically initiated a purge cycle

Wait for the purge cycle to clear

the blockage and the module to

be reset.

Check calibration gas

concentration.

The calibration gas is not flowing

Check that the calibration gas is

turned on.

The calibration gas has the wrong CO2

concentration

Check that the calibration gas

has the correct concentration.

Check calibration gas flow.

The calibration gas flow is unstable

Check the sampling line for kinks

or obstructions.

Check all tubing connections for

leaks.