beautypg.com

Specifications, Physical specifications – Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual

Page 201

background image

Specifications

Physical specifications

Protection classifications, all monitor configurations

Characteristic

Specification

Electrical rating

100 – 240 V AC, 50 – 60 Hz, 0.8 –1.5 A

Duty cycle

Continuous operation

Type of protection against electric shock

Class I equipment (protectively earthed) with double insulation

Degree of protection against electric

shock, for parts applied to patients

Type BF defibrillator proof
IEC EN 60601-1, 2nd Edition

Recovery time following defibrillator

discharge

Less than or equal to 10 seconds

Flammable anesthetics

WARNING Not suitable for use with flammable
anesthetics.

Degree of protection provided by the

enclosure with respect to harmful ingress

of liquids

IPX0
Non-protected according to EN/IEC 60529; Pulse oximeter equipment

complies with ISO 9919 Cl. 44.6 Ingress of liquids tests and EN/IEC 60601-1,

60601-2-30, 60601-2-49 Cl. 44.3 Spillage tests

Height

Standard chassis: 10 in. (25.4 cm)
Extended chassis: 10 in. (25.4 cm)

Width

Standard chassis: 11 in. (27.9 cm)
Extended chassis: 11.38 in. (28.9 cm)

Depth

Standard chassis: 6 in. (15.3 cm)
Extended chassis: 7.5 in. (19.1 cm)

Weight (including battery)

Standard chassis: 9.5 lb (4.3 kg)
Extended chassis: 10.4 lb (4.7 kg)

195