The monitor is not connected to a power outlet and is running on battery power. The

estimated charge time remaining is displayed in the hour(s):minute(s) format. Each

section of the battery status indicator represents a percentage of remaining charge.

Note

When the battery switches from AC power to battery power,
no charge time is displayed while the device calculates the
remaining charge time.

The monitor is connected to a power outlet but the battery does not maintain a

charge or has been removed.

When the battery is not being recharged and power becomes low, a very low-priority

alarm appears in the Device Status area.

Note

Observe the remaining battery charge in the battery status indicator and
plug the monitor into a power outlet as soon as you are able.

If the very low-priority alarm is dismissed or if you take no action to charge the battery, a

high-priority alarm appears when battery power is critically low. Plug the monitor into a

power outlet immediately to prevent it from powering down.

Failure to plug in the monitor before the remaining charge runs out causes the monitor to

power down and lose all unsaved data.

Alarm and information messages

The Device Status area provides alarm and information messages that are either

temporary or exist as long as the condition to which the message applies remains. Alarm

or information messages may also include controls and/or behavior that you can use to

manage alarm and information messages.

When the monitor detects an alarm condition, an alarm message appears. When

multiple alarms occur, the device provides options to manually toggle or autoscroll

through a stack of messages. You can cycle through each alarm message in sequence of

priority by touching the multiple alarm toggle. Alternatively, the autoscroll option displays

messages in the stack for approximately 4 seconds. These messages are stacked in

order of priority: the monitor displays the highest priority messages first followed by

each successive alarm message in order of priority. After presenting the entire stack of

alarm messages, the autoscroll feature loops back to the top priority message and

repeats the entire stack unless specific alarms auto-dismiss or you reset or cancel

specific alarms. In both the manual toggle and autoscroll options, when multiple alarm

messages of the same priority exist at the same time, the device displays the most

recent of those alarm messages first.

Information messages instruct you to interact with the monitor in a specific way or

provide information that does not require action. You can dismiss an information