Reset the monitor, Set the date and time, Enter clinician information – Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual

Page 43

background image

Power down.

When the Continuous Monitoring profile is not active: The device performs a

complete software shutdown, and any patient context and patient measurement

data displayed onscreen are cleared.

When the Continuous Monitoring profile is active, but no trend data has been

stored: The device performs a complete software shutdown.

When the Continuous Monitoring profile is active and trend data has been

stored:

If you opted to save patient data, existing patient context (if established) and

trend data are stored and can be re-established on the next power up.

If you opted to save patient data but no patient context was established,

trend data is stored under a temporary ID and can be re-established on the

next power up.

If you opted to remove patient data from the device, monitoring and patient

assignment end, and trend data on the device is deleted.

Cancel. This button dismisses the dialog and returns you to the previous screen.

Note

If your monitor is configured with a Braun thermometer, avoid powering
down the monitor between patient measurements to help maintain an
adequate charge to the Braun dock.

Reset the monitor

If the monitor stops functioning, you can press and hold

for approximately 6

seconds to allow the hardware to completely cycle off and to reset the monitor

configuration settings to the last saved default power-up configuration. The button is

located on the side of the monitor.

CAUTION Do not use a long press of

to power down the monitor

when it is functioning normally. You will lose patient data and configuration
settings.

Set the date and time

1. Touch the Settings tab.
2. Touch the Device tab.
3. Touch the Date/Time vertical tab.

4. To change the date and time values: Touch the up and down arrow keys or touch

and enter a value.
Repeat for each value you want to change.

Note

The date and time stamps on saved patient measurements
will adjust in response to new date and time settings.

Enter clinician information

1. Go to the Clinician tab using one of these methods:

Directions for use

Startup 37