Patient rest mode – Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual
Page 91

Icons in parameter frames
Icon
Name and status
Alarm audio paused.
The audio tone is paused. The icon remains until the paused time counts down to 0.
Icons in the Device Status area
The icons in the Device Status area are black and white, but the background area
changes colors to indicate the alarm priority. Messages accompany these icons. These
icons can be controls or status indicators.
Icons in the Device Status area
Icon
Name and status
Alarm active.
One or more alarms are active. Touch this icon to reset (pause or turn off) the alarm tone.
WARNING Do not pause or turn off an audible alarm if patient safety might be
compromised.
Alarm audio off.
Audio signals are disabled, but alarm limits and visual alarm signals remain active.
If the patient is being monitored continuously, maintain a clear line of sight with the monitor while
audio alarms are disabled.
Multiple alarms toggle.
Touch this icon to cycle through the messages for each active alarm.
Alarm audio paused.
The audio tone is paused. The icon remains until the paused time counts down to 0. Touch this
icon to start the user-configured pause interval (set in the Advanced tab).
Patient rest mode
Patient rest mode enables you to turn the audio off and dim the screen of the device
when continuous monitoring is enabled and a device is connected to the Central Station.
The visual alarms indication on the device will display, and audible alarms will sound at a
compatible Welch Allyn host with visual indications. The device may still be used for
additional vitals capture. Touch the screen to remove the dimming while in use. If a loss
of connectivity occurs, certain alarms become audible until connectivity is restored.
Patient rest mode can be enabled and disabled from the Central Station or from the
device. If patient rest mode is initiated from the Central Station, the mode can be
disabled at the Central Station or at the device. If patient rest mode is initiated from the
device, it must be disabled at the device.
Directions for use
Alarms 85