Set up NIBP

Follow these steps to configure NIBP parameter settings and to set up modifiers.

1. Touch the Settings tab.
2. Touch the Setup tab.
3. Touch the NIBP vertical tab.
4. Adjust the following setting as desired:

Cuff inflation target. If the monitor is set up for StepBP, enter the Cuff inflation

target. You can set the default in Advanced settings.

5. Touch the Home tab.

The new setting takes effect immediately.

6. To set up modifiers, touch the Patients tab.
7. Touch the Manual tab.
8. Scroll through the list to find the NIBP section, then enter or select the NIBP

modifier(s) as desired.

Cuff site. Select the site from the list box.

Cuff size. Select the size from the list box.

Patient position. Select the position from the list box.

9. Enter or select other modifier(s), if applicable, as desired.
10. Touch OK.

The Home tab appears. In the Continuous Monitoring profile, the modifiers are saved

with the next set of measurements sent to the network. In the episodic profiles, the

modifiers are saved with the next set of measurements you take or with any current,

unsaved measurements on the device when you touch Save.

Configure NIBP alarms

Follow these steps to set alarm limits for systolic and diastolic measurements, and MAP

calculation.

1. Verify that you are using the Intervals Monitoring or Continuous Monitoring profile.
2. Touch the Alarms tab.
3. Touch the NIBP vertical tab.
4. Ensure that the NIBP and MAP alarm limit controls are set to ON.

Note

If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals will occur for that specific parameter.

5. Adjust the following settings as desired:

Systolic. Enter upper and lower Systolic alarm limits using the up/down arrow

keys or the keypad.

Diastolic. Enter upper and lower Diastolic alarm limits using the up/down arrow

keys or the keypad.

MAP. Enter upper and lower MAP alarm limits using the up/down arrow keys or

the keypad.

Directions for use

Patient monitoring 125