NIBP specifications
Pediatric: 280 mmHg (37.3 kPa) (StepBP, SureBP)
Neonate: 130 mmHg (17.3 kPa) (StepBP)
Blood pressure determination time
Typical: 15 seconds
Maximum: 150 seconds
Blood pressure accuracy
Meets or exceeds ANSI/AAMI SP10:2002 standards for noninvasive blood
pressure accuracy, ±5 mmHg (0.7 kPa) mean error, 8 mmHg (1.1 kPa)
standard deviation
Mean Arterial Pressure (MAP) range
The formula used to calculate MAP yields
an approximate value.
Adult: 23 to 230 mmHg (3.1 to 30.7 kPa) (StepBP, SureBP)
Pediatric: 23 to 230 mmHg (3.1 to 30.7 kPa) (StepBP, SureBP)
Neonate: 13 to 110 mmHg (1.7 to 14.7 kPa) (StepBP)
Pulse rate range (using blood pressure
determination)
Adult: 30 to 200 beats per minute (StepBP, SureBP)
Pediatric: 30 to 200 beats per minute (StepBP, SureBP)
Neonate: 35 to 220 beats per minute (StepBP)
Pulse rate accuracy (using blood pressure
determination)
±5.0% (±3 beats per minute)
Overpressure cutoff
Adult: 300 mmHg ±15 mmHg (40.0 kPa ±2.0 kPa)
Pediatric: 300 mmHg ±15 mmHg (40.0 kPa ±2.0 kPa)
Neonate: 150 mmHg (20.0 kPa) maximum
SureTemp Plus temperature module specifications
Units of measure
°F, °C; user-selectable
Temperature range
80°F to 110°F (26.7°C to 43.3°C)
Calibration accuracy
±0.2°F (±0.1°C) (Direct mode)
Braun ThermoScan PRO 4000 thermometer specifications (refer to manufacturer's directions for
use for additional information)
Units of measure
°F, °C; user-selectable
Temperature range
68°F to 108°F (20°C to 42.2°C)
Calibration accuracy
•
±0.4°F (±0.2°C) for temperatures ranging from 95.9°F to 107.6°F
(35.5°C to 42°C)
•
±0.5°F (±0.3°C) for temperatures outside of this range
Display resolution
0.1°F or °C
198 Specifications
Welch Allyn Connex
®