Dynamic range, Third order intermodulation distortion, Noise density at 1 ghz – Agilent Technologies N9010A User Manual
Page 87: Residual responses, Image responses, Lo related spurious, Other spurious
Chapter 3
87
VXA Vector Signal and WLAN Modulation Analysis Application
Vector Signal Analysis Performance (N9064A-1FP/1TP)
Dynamic Range
Description
Specs & Nominals
Supplemental
Information
Third Order Intermodulation
distortion
−84 dBc (nominal)
(Two
−20 dBFS tones,
400 MHz to 13.6 GHz,
tone separation > 5
× IF Prefilter
BW)
Noise Density at 1 GHz
Input Range
≥−10 dBm
−137 dBFS/Hz
−20 dBm to −12 dBm
−127 dBFS/Hz
−30 dBm to −22 dBm
−129 dBFS/Hz
requires preamp option
−40 dBm to −32 dBm
−119 dBFS/Hz
requires preamp option
Residual Responses
−90 dBFS (nominal)
(Range
≥ −10 dBm)
Image Responses
−75 dBc
(10 MHz to 13.6 GHz,
<8 MHz span)
LO Related Spurious
−60 dBc
(10 MHz to 3.6 GHz,
f > 600 MHz from carrier)
Other Spurious
(<8 MHz span)
100 Hz < f < 10 MHz from
carrier
−70 dBc (nominal)
f
≥ 10 MHz from carrier
−70 dBc
−70 dBc (nominal)