Lo output – Agilent Technologies N9010A User Manual
Page 134
134
Chapter 11
Option EXM - External Mixing
Other External Mixing Specifications
c. The amplitude accuracy of a measurement includes this term and the accuracy with which the settings of
corrections model the loss of the external mixer.
Description
Specifications
Supplemental
Information
LO Output
Frequency Range
3.75 to 14.0 GHz
Output Power
a
20 to 30
°C
Full Range
3.75 to 7.0 GHz
b
+15.0 to 18.0 dBm
+14.5 to 18.5 dBm
7.0 to 8.72 GHz
+15.0 to 18.0 dBm
+13.5 to 18.8 dBm
7.8 to 14.0 GHz
+14.0 to 18.5 dBm
Not specified
Second Harmonic
–20 dB (nominal)
Fundamental Feedthrough
and Undesired Harmonics
c
–15 dB (nominal)
VSWR
<2.2:1 (nominal)
a. The LO output port power is compatible with Agilent M1970 and 11970 Series mixers except for the
11970K. The power is specified at the connector. Cable loss will affect the power available at the
mixer. With non-Agilent mixer units, supplied loss calibration data may be valid only at a specified LO
power that may differ from the power available at the mixer. In such cases, additional uncertainties
apply.
b. LO Doubler = Off settings.
c. LO Doubler = On setting. Fundamental frequency = 3.9 to 7.0 GHz.