Qpsk evm – Agilent Technologies N9010A User Manual

Page 294

background image

294

Chapter 33

W-CDMA Measurement Application

Measurements

Description

Specifications

Supplemental Information

QPSK EVM

(

−25 dBm ≤ ML

a

≤ −15 dBm

20 to 30°C)

RF input power and attenuation are set to
meet the Mixer Level range.

EVM

Range

0 to 25% (nominal)

Floor

1.6%

Accuracy

b

±1.0%

I/Q origin offset

DUT Maximum Offset

−10 dBc (nominal)

Analyzer Noise Floor

−50 dBc (nominal)

Frequency error

Range

±30 kHz (nominal)

c

Accuracy

±5 Hz + tfa

d

a. ML (mixer level) is RF input power minus attenuation.
b. The accuracy specification applies when the EVM to be measured is well above the measurement floor

and sucessfully synchronized to the signal. When the EVM does not greatly exceed the floor, the errors
due to the floor add to the accuracy errors. The errors due to the floor are noise-like and add incoher-
ently with the UUT EVM. The errors depend on the EVM of the UUT and the floor as follows: error =

sqrt(EVMUUT

2

+ EVMsa

2

)

− EVMUUT, where EVMUUT is the EVM of the UUT in percent, and

EVMsa is the EVM floor of the analyzer in percent.

c. This specifies a synchronization range with CPICH for CPICH only signal.
d. tfa = transmitter frequency

× frequency reference accuracy