28

Chapter 1

Agilent EXA Signal Analyzer

Frequency and Time

External Triggers

See

“Trigger Inputs” on page 72

TV Triggers

Triggers on the leading edge of the selected
sync pulse of standardized TV signals.

Amplitude Requirements

–65 dBm minimum video carrier power at
the input mixer, nominal

Compatible Standards

NTSC-M,
NTSC-Japan,
NTSC-4.43,

PAL-M, PAL-N,
PAL-N Combination,
PAL-B/-D/-G/-H/-I.
PAL-60, SECAM-L

Field Selection

Entire Frame, Field
One, Field Two

Line Selection

1 to 525, or 1 to 625,
standard dependent

a. The highest allowed mixer level depends on the IF Gain. It is nominally –10 dBm for Preamp Off and

IF Gain = Low.

b. Noise will limit trigger level range at high frequencies, such as above 15 GHz.

Description

Specifications

Supplemental Information

Gated Sweep

Gate Methods

Gated LO

Gated Video

Gated FFT

Span Range

Any span

Gate Delay Range

0 to 100.0 s

Gate Delay Settability

4 digits,

≥100 ns

Gate Delay Jitter

33.3 ns p-p (nominal)

Gate Length Range

(Except Method = FFT)

100 ns to 5.0 s

Gate length for the FFT method is fixed at
1.83/RBW, with nominally 2% tolerance.

Gated Frequency and
Amplitude Errors

Nominally no additional error for gated
measurements when the Gate Delay is
greater than the MIN FAST setting

Gate Sources

External 1

External 2

Line

RF Burst

Periodic

Pos or neg edge triggered

Description

Specifications

Supplemental Information