Frequency response, Absolute amplitude accuracy, Frequency response absolute amplitude accuracy – Agilent Technologies N9010A User Manual
Page 119
![background image](/manuals/35678/119/background.png)
Chapter 8
119
Option EA3 - Electronic Attenuator, 3.6 GHz
Other Electronic Attenuator Specifications
Description
Specifications
Supplemental Information
Frequency Response
Mech atten set to
default/calibrated setting of
10 dB.
(Maximum error relative to
reference condition (50 MHz))
20 to 30
°C
Full Range
95th Percentile (
≈2σ)
Attenuation = 4 to 24 dB, even
steps
9 kHz to 10 MHz
±0.75 dB
±0.90 dB
±0.32 dB
10 MHz to 50 MHz
±0.65 dB
±0.69 dB
±0.27 dB
Option
≤
526
50 MHz to 2.2 GHz
±0.48 dB
±0.60 dB
±0.19 dB
2.2 to 3.6 GHz
±0.55 dB
±0.67 dB
±0.20 dB
Option
>
526
50 MHz to 2.2 GHz
±0.48 dB
±0.70 dB
±0.19 dB
2.2 to 3.6 GHz
±0.55 dB
±0.70 dB
±0.22 dB
Attenuation = 0, 1, 2 and odd
steps, 3 to 23 dB
10 MHz to 3.6 GHz
±0.30 dB
Description
Specifications
Supplemental Information
Absolute Amplitude Accuracy
At 50 MHz
a
20 to 30
°C
Full temperature range
±0.44 dB
±0.47 dB
At all frequencies
20 to 30
°C
Full temperature range
±(0.44 dB + frequency response)
±(0.47 dB + frequency response)
a. Absolute amplitude accuracy is the total of all amplitude measurement errors, and applies over the fol-
lowing subset of settings and conditions: 1 Hz
≤ RBW ≤ 1 MHz; Input signal −10 to −50 dBm; Input
attenuation 10 dB; all settings auto-coupled except Swp Time Rules = Accuracy; combinations of low
signal level and wide RBW use VBW
≤ 30 kHz to reduce noise. When using FFT sweeps, the signal
must be at the center frequency.
This absolute amplitude accuracy specification includes the sum of the following individual specifica-
tions under the conditions listed above: Scale Fidelity, Reference Level Accuracy, Display Scale
Switching Uncertainty, Resolution Bandwidth Switching Uncertainty, 50 MHz Amplitude Reference
Accuracy, and the accuracy with which the instrument aligns its internal gains to the 50 MHz Ampli-
tude Reference.